Comments
The 2nd error should be fixed in tomorrow's package. (PlusToolkit/PlusApp@9064af0) I'm compiling a new build to test the crash. Can you upload your config file? |
Thank you so much. PlusDeviceSet_fCal_Sim_TemporalCalibration - Clarius.txt In the config file apart from StylusTip to Stylus Transform (which I computed), other matrices are the matrices in the example file. I also wanted to ask is it required to change them? |
Hello, |
I can't replicate the crash with temporal calibration. Can you confirm if it works when you load the recorded temporal simulation that comes with fCal? |
Yes, when I load "PlusDeviceSet_fCal_Sim_TemporalCalibration.xml" it works. However, when I loaded my config file (for the Clarius device) and I wanted to open Temporal calibration it crashed. |
Hello, After a while of checking and changing parameters in the config file for temporal calibration, I figure out that when I set the “PortUsImageOrientation” to “FM” I am able to open the Temporal Calibration tab without a crash. However, when I open Temporal Calibration in fCal.exe, I get the below errors: I appreciate it if you can help to solve this issue. |
Hello @Sunderlandkyl, Would you please let me know if there is any way to solve this issue? |
Can you set the log level to Trace in fCal and attach a log file from the crash? |
Yes, this is the log file when PortUsImageOrientation="MF" and the crash happens. Below is the log file when PortUsImageOrientation="FM" and I get errors. |
The first log doesn't seem to provide info into the cause of the crash. Do you have experience in compiling a debug build to test where the crash occurs? Otherwise I'll see if I can get access to a Clarius or find someone with a debug build. |
Hello @Sunderlandkyl , |
Hello, |
Hello @Sunderlandkyl, |
I'll have access to the device again soon. I'll take a look at it then. |
I really appreciate it. |
I've managed to reproduce the crash. Looking into it. |
Thank you so much. |
Yes actually, it looks like it's due to the fact that the image is not grayscale. |
Thank you so much. |
Hello,
I am a new user of 3d Slicer and Plus and I want to do the volume reconstruction. As I understand, first, I should do the “tracked ultrasound calibration”. Therefore, now I am trying to calibrate the Clarius ultrasound probe with a tracker device (OptiTrack).
1.I am able to do the “Pivot calibration” By using SlicerIGT Tutorial U_11 or using fCal.exe from “PlusApp-2.9.0.20220210-Clarius-Win64”.
2 . After that I wanted to do the temporal calibration using fCal.exe. I used “PlusApp-2.9.0.20220210-Clarius-Win64”. I created a fCal configuration file. However, when I try to open “Temporal calibration” fCal crashes and exits.
3 . Also, I tried to use fCal.exe from “PlusApp-2.9.0.20220304-Clarius-Win64”. However, with this version I am not even able to do the “Pivot calibration” it shows “Stylus calibration configuration is missing!” and shows below errors:
|ERROR|036.537000| Unable to find required vtkIGSIOPivotCalibrationAlgo element in device set configuration| in E:\D\PTNPCb\IGSIO\IGSIOCalibration\vtkIGSIOPivotCalibrationAlgo.cxx(286)
|ERROR|036.537000| Reading pivot calibration algorithm configuration failed!| in E:\D\PTNPCb\PlusApp\fCal\Toolboxes\QStylusCalibrationToolbox.cxx(68)
For “Temporal calibration” in this version it also crashes when I click on “Temporal calibration”.
Can you please help me in this regard?
Thank you.