SECoP Issue 56: Additional Busy States (closed)

Motivation

This issue is raised when discussion the possible substates of a cryomagnet with persistency. After reaching the target field, in principle a measurement can start, but the action still has to continue, running down the current in the leads. The question arised, how to deal with this. Is this BUSY or IDLE? Should we introduce a new substate for this?

Proposal

The proposal of Klaus is, to introduce FINALIZE, which is a substate of BUSY. If the ECS sees that a module has this state, it might offer the user the possibility to choose wheter to wait or not for this state to be finished.

Discussion

Markus proposed, that this should be handled differently. If the SEC-Node programmer wants to offer the possibility to choose, (s)he implements a parameter named like "finalize_is_busy", which lets the user choose. Actually, such situations are handled differently at different facilities: At MLZ, measuring contimues only after finalizing, at PSI/SINQ, the measurement continues and ramping down the leads is done in background.

Markus thinks, that a more general discussion of the BUSY state has to be done before this decision. BUSY may have different meanings:

  1. waiting for value to reach target
  2. an action is not finished
  3. we have to wait before we can do the next action

Markus sees the BUSY status mainly with meaning (1). Meaning (2) seems not relevant for the ECS: as long as the SEC-Node can handle to accept new actions while an other is still running, nobody has to care. As there are SEC node developers which allow to change the target while BUSY, meaning (3) can also not be the main maining of BUSY. We discussed that when an action is started while something is still running, an the SEC-node can not handle this, it must report an error message with the error class "IsBusy".

What we did not define, is, if BUSY has only the meaning (1) or if it is a logical OR of (1) and (3). In the first case, an "IsBusy" error might be the reply of an action (change target or command), even when the status is IDLE.

If we want be very clear with this, we may need an additional boolean parameter "running", which only tells that the value on the way to the target. status=BUSY would then have meaning (3).

If the committee does not like coming back to the solution of the "finalize_is_busy" parameter, Markus proposal for a compromise would be:

  • status=BUSY is logical OR of meaning (1) and (3)
  • we leave it open to the SEC node developer to offer a "finalize_is_busy" parameter or to use a newly defined substate EARLY_IDLE=101 (substate of IDLE!).
EARLY_IDLE means then:
target is reached AND ready to accept an other "change target".
We may in addition introduce FINALIZE=301 (substate of BUSY!), which means:
target is reached, measurement might continue, but 'change target' is not yet accepted.

Additional considerations after vidconf-2019-04-11

Use cases wich may need substates

In the following, use cases were collected, which might need substates or additional commands, but let us start with the simplest use case:

A) Simple use case: Wait until a value has reached target

This is the simplest use case, and as it is probably also the main usage, it must be simple: After a change target (or go), the ECS has to wait before the value has reached target, before it is collecting some data under stable conditions. For this use case, the simplest solution is, that the transition from BUSY to IDLE happens when stable conditions are reached, and in the most efficient solution, this happens as soon as possible, not waiting for example for cooling the switch heater and ramping down the current in the leads of a superconducting magnet.

For the example of the superconducting magnet, an implementor might choose to stay BUSY while ramping down the leads for simplicity - this is o.k., it is just not ideal, because it is a waste of time.

B) Preparing

For special cases, measurements might already happen while ramping the main value. For this purpose, the 'prepare' command is proposed. The meaning is: bring the module into a state, where the 'moving' of the main value can start immediately. It seems not necessary to introduce an additional state PREPARING for that: The status would be BUSY while preparing, IDLE afterwards. However, for being explicit, we might introduce PREPARED, which would be a substate of IDLE. This additional state is needed, if we define 'IDLE' as: 'the value is at target'.

C) Actions affecting the main value (Example: reference run)

In addition to the (A) and (B), there might be other actions, which we want to know when they end. There are different types of such actions. For example a reference run on a motor: during a reference run, the value is obviously not at target, so it is clear, that after a command 'reference_run' the module is BUSY until the reference run is finished, and target and value should be identical after a reference run.

D) Actions not affecting the main value (Example: going persistent)

Going from persistent mode to non persistent mode is an example for an action, which does not influence the main value. The user might want to do this in order to save time on the next field change. Or (s)he want to do the inverse to save helium. The question is: is it is necessary for the ECS to wait for the end of such an action? The status text might give an indication of the progress of such an action to the user, but the ECS will normally not care. With the exception of the case, where the SEC-Node software does not accept new actions while others are running. This is handled by the next use case.

E) Actions blocking new commands

If the ECS wants to initiate an action, but the SEC-Node can not yet accept it, because it is still busy with an other action, the reply would be an 'IsBusy' error message. The question is then: when does it make sense for the ECS to retry? A sensible answer is: whenever the status changes or more precise: whenever the status code or the status text changes. This means that we do not need substates for this meaning, it is enough for the SEC-node to change the status text. It is clear, that the ECS may get an IsBusy error again, but it is not forced to try repeatedly and guess what frequency is meaningful. Above is valid for asynchronous communication. In case of synchronous operation, anyway a polling has to be done. Whether the client polls the status, or retries the operation comes more or less to the same.

With this in mind, a defined BUSY substate with the meaning (3) is not really needed. Actions which do not affect the main value, but can not be interrupted, may just be IDLE with a different status text, or may be a custom substate of IDLE. However, if the committee decides, that it is better to use a defined substate for that, we might define new substate of IDLE, with the meaning (3) of busy. However a name BUSY_IDLE or IDLE_BUSY seems not very nice. In this case the name of the substate should be chosen by the implementor, for example CHANGING_PERSISTENCY.

F) Influencing the end of the BUSY phase

Some users may want to influence the criterium for the end of BUSY phase.

For example:

  1. During cooldown of the superconducting switch, the magnetic field might still oscillate slightly, so the user wants to wait for this before measuring.
  2. Not really a sample environment issue, but otherwise a good example: the user wants to wait until air cushions have switched off, because the beam geometry is affected sligthly.
  3. Tolerance and window of temperature

Instead of additional substates, the SEC-node may offer one or several additional parameter(s), influencing, when the transition to IDLE happens. This is already the case in example (3) above. For the example of motors with air cushions, this might be a parameter settling_time, defining how long to wait after the air cushion was switch off. Or, for a magnet, it might be a parameter "complete_on" with the values "field_at_target", "switch_closed" and "leads_at_zero".

The naming and meaning of these parameters may be defined for specific interface classes, but otherwise we should not try to find a more generic meaning.

The advantage of this approach is, that it is "self documented", by the selection of the parameter names, and the description of the parameter.

A disadvantage is, that the criteria are preselected and then valid for all clients, they can not be chosen on the fly for different clients. But do we need that really?

Decision

Useful statuscodes
code name generic variant name Meaning
0 DISABLED DISABLED Generic Module is not enabled
100 IDLE IDLE Generic Module is not performing any action
130 STANDBY IDLE Standby Stable, steady state, needs some preparation steps, before a target change is effective
150 PREPARED IDLE Prepared Ready for immediate target change
200 WARN WARN Generic The same as IDLE, but something may not be alright, though it is not a problem (yet)
230 WARN_STANDBY WARN Standby -''-
250 WARN_PREPARED WARN Prepared -''-
300 BUSY BUSY Generic Module is performing some action
310 DISABLING BUSY Disabling Intermediate state: Standby -> DISABLING -> Disabled
320 INITIALIZING BUSY Initializing Intermediate state: Disabled -> INITIALIZING -> Standby
340 PREPARING BUSY Preparing Intermediate state: Standby -> PREPARING -> PREPARED
360 STARTING BUSY Starting Target has changed, but continuous change has not yet started
370 RAMPING BUSY Ramping Continuous change, which might be used for measuring
380 STABILIZING BUSY Stabilizing Continuous change has ended, but target value is not yet reached
390 FINALIZING BUSY Finalizing Value has reached the target and any leftover cleanup operation is in progress. If the ECS is waiting for the value of this module beeing stable at target, it can continue.
400 ERROR ERROR Generic An Error occured, Module is in an error state, something turned out to be a problem.
430 ERROR_STANDBY ERROR Standby An Error occured, Module is still in Standby state, even after clear_errors.
450 ERROR_PREPARED ERROR Prepared An Error occured, Module is still in PREPARED state, even after clear_errors.